Autoclave Operational Qualification Procedure: Step-by-Step OQ Protocol

Autoclave Operational Qualification Procedure verifies that an autoclave operates reliably within its specified parameters. This covers the OQ process, operational tests, acceptance criteria, and GMP documentation required for pharmaceutical microbiology laboratories.

Read Also => Installation Qualification for Autoclave – complete procedure

Objectives of the Autoclave Operational Qualification Procedure

The objective of Operational Qualification (OQ) is to verify and document that the Vertical Autoclave – 100/300/600 Liter operates as intended throughout all defined operating ranges such as temperature, pressure, and time cycles, in accordance with manufacturer specifications and GMP requirements.

To verify that the equipment operates in accordance with the design and user requirements as defined by set Acceptance Criteria and complies with relevant cGMP Requirements.

 To verify the Operational features of Vertical Autoclave and to ensure that it produces desired Quality & rated output according to manufactures specifications.

To verify all the Operational features from user point of view of the Equipment, Cleaning Procedure, Start up & Shut down Procedure and Safety Features.

Scope where applicable

The scope of this operational qualification protocol cum report is limited to qualification of Vertical Autoclave – 100/300/600 Liter installed in between material receiving area & Sterilization area of Microbiology Section.

This Protocol will define the methods and documentation used to perform OQ activity the Vertical Autoclave for OQ. Successful completion of this Protocol will verify that Vertical Autoclave meets all acceptance criteria and ready for Performance Qualification.

Pre Approval Of Documents

Prepared by

DesignationNameDateSignature
Executive/Officer Engineering   

Reviewed By

DesignationNameDateSignature
Head Engineering   
Head Warehouse   
Head Quality Control   
Head Production   

Approved by

DesignationNameDateSignature
Head QA   

Scope of supply:

DescriptionSpecification
Make 
Model 
Control system230 V AC  Single Phase

Brief process description containing Autoclave Operational Qualification Procedure

The vertical autoclave sterilizes materials by generating saturated steam under pressure at preset temperature and time cycles (e.g., 121°C for 15–30 minutes).
OQ ensures that the control panel, sensors, and safety systems function properly during operation.

The Vertical Autoclave is used for sterilization of culture media, glassware, instruments, garments, and laboratory waste by applying saturated steam under pressure at predefined temperature and time cycles.

The process ensures effective destruction of microorganisms including bacterial spores to maintain contamination control in when the material is put inside pass box & the door is closed, UV light will turn on and material is placed under UV for some time to kill bacteria. Then the material can be taken out

Material of Construction

ComponentMaterial
Sterilization ChamberSS 316L
JacketSS 304 / SS 316
Outer BodySS 304 Matt Finish
DoorSS 316L
PipingSS316L
FrameSS304
GasketSilicone / EPDM (Food Grade)

Chamber Specification

  • Double wall construction
  • Jacketed chamber
  • Mirror polished internal surface (Ra ≤0.8 µm)
  • Rounded corners for easy cleaning
  • Complete condensate drainage
  • Self-draining design
  • No dead legs

Sterilization Parameters

ParameterSpecification
Sterilization Temperature121°C & 134°C
Working Pressure15–30 psi
Maximum Temperature140°C
Cycle TimeProgrammable
Holding Time1–99 Minutes
Drying CycleAutomatic
Cooling CycleAutomatic

Control System

  • PLC based controller
  • 7–10 inch HMI Touch Screen
  • Password protected access
  • Recipe management
  • User level access
  • Alarm history
  • Cycle history
  • Data logging
  • Real-time graphical display
  • USB data export
  • Ethernet communication
  • Optional SCADA connectivity

Instrumentation

InstrumentRequirementObservation
Temperature SensorPT100 RTD 
Chamber Pressure GaugeDigital 
Jacket Pressure GaugeDigital 
Safety ValveYes 
Vacuum SensorIf applicable 
Steam Pressure IndicatorYes 
Door Interlock SensorYes 

Utility Required        

UtilityRequirementObservation
Power Supply415 V AC ±10%, 50 Hz, 3 Phase 
SteamPlant Steam / Built-in Steam Generator 
WaterPurified Water / Soft Water 
DrainFloor Drain Connection 
Compressed Air6–8 bar (if pneumatic door) 
Cooling WaterIf required 

Instrumentation Required

InstrumentRequirementObservation
Temperature SensorPT100 RTD 
Chamber Pressure GaugeDigital 
Jacket Pressure GaugeDigital 
Safety ValveYes 
Vacuum SensorIf applicable 
Steam Pressure IndicatorYes 
Door Interlock SensorYes 

Autoclave Operational Qualification Procedure:

Before initiating OQ, verify that:

  • Installation Qualification (IQ) has been successfully completed and approved.
  • All utilities (electrical power, steam, water, compressed air, and drain) are available and within specified limits.
  • Calibration certificates for all critical instruments are valid.
  • Standard Operating Procedures (SOPs) are approved and available.
  • Preventive maintenance requirements have been completed.
  • Validation master plan approval has been obtained.
  • Test instruments used during qualification are calibrated and traceable to national standards.

 Functional Description

DescriptionFunction
ChamberProvides the sterilization space where media, glassware, instruments, and materials are placed and exposed to steam.
Lid / Door with GasketEnsures airtight sealing to prevent steam leakage and maintain required pressure.
Heater ElementHeats the water to generate steam for sterilization.
Temperature Sensor (PT-100)Measures chamber temperature and sends signal to the controller.
Pressure GaugeDisplays the internal chamber pressure during operation.
Digital / PID ControllerControls temperature, time, and cycle parameters of the autoclave.
Safety ValveAutomatically releases excess pressure to ensure safety.
Drain ValveUsed to drain used water or condensate from the chamber.
Basket / TraySupports and arranges sterilization materials properly inside the chamber.
  Alarm  Indicates cycle completion, faults, or abnormal operating conditions.

Autoclave Operational Qualification Procedure-Parameters verification

Power Supply Verification

    Procedure
    • Switch ON the main power supply.
    • Verify that the HMI/PLC powers up correctly.
    • Check for abnormal indications or error messages.
    Acceptance Criteria
    • Equipment powers up without errors.
    • Control panel functions normally.

    HMI/PLC functional verification

    Procedure
    • Verify touch screen operation.
    • Check menu navigation.
    • Verify user login functions.
    • Verify password levels.
    • Check date and time settings.
    Acceptance Criteria
    • All functions operate correctly.
    • User access levels function as specified.

    Temperature control verification

    Procedure
    • Select a sterilization cycle.
    • Set temperature to 121°C.
    • Start the cycle.
    • Observe chamber temperature during heating.

    Repeat the test at 134°C.

    Pressure Control Verification

    Procedure
    • Run a sterilization cycle.
    • Observe chamber pressure.
    • Compare displayed pressure with calibrated reference gauge.
    Acceptance Criteria
    • Pressure corresponds to sterilization temperature.
    • Pressure variation within manufacturer’s specification.

    Cycle Operation Verification

    Verify operation of each programmed cycle.

    Typical cycles include:

    • Glassware
    • Media
    • Instruments
    • Liquids
    • Waste
    • User Defined Cycle
    Acceptance Criteria
    • Selected cycle executes correctly.
    • All programmed stages complete successfully.

    Timer Verification

    Procedure
    • Set holding time (e.g., 15 minutes).
    • Compare displayed time with calibrated stopwatch.
    Acceptance Criteria
    • Timer accuracy within ±1 minute.

    Steam Generation Verification

    Procedure

    For autoclaves with built-in steam generators:

    • Fill water reservoir.
    • Start steam generation.
    • Observe steam pressure.
    Acceptance Criteria
    • Steam generated without interruption.
    • Steam pressure remains stable.

    Door Interlock Verification

    Procedure

    Attempt to open the door:

    • During heating
    • During sterilization
    • Under pressure
    Acceptance Criteria
    • Door shall remain locked.
    • Door opens only after chamber pressure reaches zero.

    Chamber Leak Test

    Procedure
    • Pressurize chamber.
    • Observe pressure for specified holding period.
    Acceptance Criteria
    • No visible leakage.
    • Pressure loss within manufacturer’s specification.

    Alarm Verification

      Verify each alarm individually.

      AlarmExpected Result
      High TemperatureAlarm generated
      High PressureAlarm generated
      Low Water LevelAlarm generated
      Door OpenAlarm generated
      Sensor FailureAlarm generated
      Cycle FailureAlarm generated
      Power FailureAlarm generated
      Acceptance Criteria
      • Correct alarm displayed.
      • Audible alarm activates.
      • Cycle response as designed.

      Component Veriifcation

      Record the results in below table.

      Component DescriptionAcceptance CriteriaObservations (Ok/Not Ok)Observed By (Sign & date)
      ChamberClean, No Damage, Proper Finish  
       Lid / Door with GasketProper Sealing, No Leakage  
      Heater ElementHeating Properly  
      Temperature Sensor (PT-100)Accurate Reading ±2°C  
      Pressure GaugeCorrect Display ±2 PSI  
      Digital / PID ControllerDisplay & Keys Working  
      Safety ValveFunctional on High Pressure  
      Drain ValveSmooth Operation, No Leakage  

      Safety Features & Alarms

      Safety featureFunctionIdentified by/Date
      Electrical InsulationSafety 
      EarthingTo avoid the shock 
      Over-TemperatureCut-off 
      AlarmAudio/Visual Alarm Emergency Power Cut Switch Low Water Level Protection Pressure Release Valve   

      Acceptance Criteria

      Operational Qualification shall be considered satisfactory if:

      • All operational tests meet predefined acceptance criteria.
      • No critical deviations remain unresolved.
      • All alarms and safety interlocks function correctly.
      • Control system performs as intended.
      • Temperature, pressure, and timing remain within specified limits.
      • Required documentation is complete and approved.

      Deviations

      Any deviation observed during Operational Qualification shall be documented, investigated and assessed.

      _________________________________________________________________________________________________________________________________________________________________________________________________________________________________

      Conclusion

      Upon successful completion of all Operational Qualification tests, the laboratory autoclave shall be considered qualified for operational use and suitable to proceed to Performance Qualification (PQ), where sterilization performance is verified using representative loads and biological indicators.

      Post Approval Of Documents

       Prepared by

      DesignationNameDateSignature
      Executive/Officer Engineering   

       Reviewed By

      DesignationNameDateSignature
      Head Engineering   
      Head Warehouse   
      Head Quality Control   
      Head Production   

      Approved by

      DesignationNameDateSignature
      Head QA   

      List Of Abbreviations

      PSIPound per Square Inch
      MCBMiniature Circuit Board
      ISOIndian Standardized Organisation
      GMPGood Manufacturing practices
      MMMillimetre
      VVoltage
      ACAlternate Current
      PIDProportional Integral Derivative

      Leave a Comment